DigestAID - Artificial Intelligence Development, S.A.
A University of Porto spin-off's answer to capsule endoscopy's core problem, tens of thousands of frames per exam and reader fatigue, Deep Capsule reached the US market in March 2026 on a narrower claim than its research lineage suggests: lesion detection and reading time only, with no lesion differentiation in the cleared software. The FDA-reviewed evidence is a 330-exam, seven-centre, four-country retrospective reader study in which AI-assisted reading found at least one clinically relevant lesion in 96.1 percent of patients against 76.1 percent for standard reading, established as non-inferior, with reading time reduced. What holds it at one mark is the shape of that evidence: retrospective, manufacturer-run, weak on specificity (75 percent standalone on eight lesion-free patients; 12.5 percent AI-assisted in the reader study), and a US market history that began this spring.
Performance Metrics
Clinical Evidence
Two bodies of evidence need to be kept apart. The first is the device evidence the FDA reviewed for K250655, summarised in the public 510(k) summary. The second is a run of peer-reviewed papers from the Porto gastroenterology and engineering group that founded DigestAID; none of those papers names Deep Capsule, and the models they describe differentiate lesion types and grade bleeding risk, which the cleared US software does not. They are background research, not performance data for the product on sale. The FDA-reviewed standalone test used 272 patients held out from a 1,133-patient, 321,357-image dataset collected at two Portuguese centres. Patient-level sensitivity was 95.8 percent (95 percent CI 92.7 to 97.9) and specificity 75 percent (95 percent CI 34.9 to 96.8), with an AUC of 0.966. The specificity figure rests on eight lesion-free patients, hence the very wide interval; the FDA summary itself notes that clinicians will need to review the findings carefully to identify false positives. Image-level sensitivity and specificity on 101,802 frames were 92.1 and 88.0 percent. The clinical validation is a retrospective multicentre reader study of 330 capsule examinations recorded between January 2021 and April 2024 at seven centres in Portugal, Spain, Brazil and the United States (118 US exams from a single site), read by 15 gastroenterologists, with a separate expert board as reference standard. Diagnostic yield, defined as the proportion of patients with at least one clinically relevant small-bowel lesion, was 96.1 percent (93.4 to 97.8) for AI-plus-physician reading against 76.1 percent (71.2 to 80.3) for standard reading and 87.9 percent for the expert board; non-inferiority was established (p below 0.001). Per-patient sensitivity rose from 76.2 to 97.2 percent, while specificity fell from 25 to 12.5 percent on 40 lesion-free patients, so the gain is bought with more positives for the reader to dismiss. The summary reports mean reading time as significantly reduced without stating the figures; the published version of the study gives a mean AI-assisted reading time of 203 seconds per examination. That published version, Mascarenhas Saraiva et al. in the American Journal of Gastroenterology (online August 2025, print April 2026), describes the same 330-exam, seven-centre, four-country cohort and reports a lesion detection rate of 96.1 versus 76.3 percent and sensitivity of 97.5 versus 78.2 percent, with minor differences from the FDA-reviewed figures. Two cautions: the abstract does not name the product and reports a model that also differentiates lesions, and it describes the design as prospective where the FDA summary calls the study retrospective. This entry follows the FDA description. The three earlier Porto papers (BMJ Open Gastroenterology 2021; Endoscopy International Open 2024; GE Portuguese Journal of Gastroenterology 2024) report frame-level convolutional-network results for lesion differentiation, Saurin bleeding-risk stratification and multi-brand vascular-lesion detection. They are listed under publications as research models. Joao Ferreira, the engineering lead, is disclosed as a paid DigestAID employee in both 2024 papers; the 2021 paper declares no competing interests and predates the company's founding. The honest gap is not the absence of a reader study but its provenance and design: every study is manufacturer-run, the validation is retrospective, independent validation of the cleared software has not been published, and no evidence addresses use outside the suspected-small-bowel-bleeding indication.
| Study | Design | n | Sensitivity | Specificity | AUC | Published |
|---|---|---|---|---|---|---|
| DigestAID clinical validation study DC001 (manufacturer-run; 15 readers, 7 centres; reviewed by FDA for K250655) | RetrospectiveRetrospective | 330 | 97.2% AI-aided vs 76.2% standard (per patient) | 12.5% AI-aided vs 25% standard (40 lesion-free patients) | — | K250655 510(k) summary, FDA, Mar 2026 — diagnostic yield 96.1% vs 76.1%; peer-reviewed counterpart Am J Gastroenterol 2026 (PMID 40874973) |
| DigestAID standalone algorithm testing (manufacturer-run; reviewed by FDA for K250655) | RetrospectiveRetrospective | 272 | 95.8% (92.7–97.9) | 75% (34.9–96.8) | 0.966 (0.939–0.993) | K250655 510(k) summary, FDA, Mar 2026 — internal test set from 2 Portuguese centres; image-level 92.1% / 88.0% on 101,802 frames |
Inside the algorithm
Editorial featureHow Deep Capsule reads a capsule study.
Five stages — from raw input to verdict — drawn from manufacturer documentation and the public regulatory record.
- INGEST
- NORMALISE
- DETECT
- LOCALISE
- VERDICT
Stage 01 · INGEST
A recorded small-bowel capsule study arrives for analysis.
A small-bowel capsule study is recorded as normal: the patient swallows the capsule, and the hours-long transit produces a video that can run to tens of thousands of frames. Deep Capsule takes the completed, previously acquired study as its input; nothing happens in real time during transit.
The cleared indication fixes the population: adults examined for suspected small-bowel bleeding, read on the compatible capsule systems named in the labelling. In the FDA-reviewed validation that meant PillCam SB3 for 93 percent of exams, with small Olympus EC-10 and PillCam SB2 subsets.
Input
Capsule-endoscopy video frames
Inside the Auris+ Listing
Five more sections complete this device’s Auris+ Listing.
Decision Ledger
ProPro unlocks a private, cross-vendor log of every case you read on this device — what the AI called, what you concluded, and a one-line reason if you overrode it. Ready for the EU AI Act's deployer logging obligations when they land in 2028.
Clinical Evidence Deep Dive
ProPro unlocks the structured clinical-evidence summary — study count, target patient population, and a tabular accuracy-metrics view drawn from peer-reviewed sources.
Peer-Reviewed Publications
ProPro unlocks the curated peer-reviewed publication list with PubMed cross-links — the citation backbone of every editorial verdict.
Post-Market & Regulatory Conditions
ProPro unlocks the post-market surveillance summary, recall record, and the conditions of approval that bound real-world use.
AI Algorithm Version History
ProPro unlocks the chronological record of algorithm version changes — what changed when, drawn from manufacturer changelogs and regulatory filings.
Regulatory Approvals
Safety Record
No FDA recall or adverse-event cluster attributable to Deep Capsule was located in public reporting as of the review date; the FDA enforcement and MAUDE databases could not be queried directly from this environment, and the device's US market history only began in March 2026, so the post-market signal is short by construction. The harm surface of the category is indirect: the software selects frames for the clinician rather than issuing a diagnosis, so the principal risk is a missed lesion in frames the reader de-prioritises, and the labelling carries an explicit warning against over-reliance. The FDA-reviewed data describe a sensitivity-first profile: patient-level standalone sensitivity of 95.8 percent, and in the reader study 8 of 290 expert-board-positive patients missed by AI-assisted reading against 69 missed by standard reading. The cost is on the other side of the ledger: standalone specificity of 75 percent on eight lesion-free patients and AI-assisted per-patient specificity of 12.5 percent mean a large share of flagged frames will be false positives that the reader must clear. Procurement reviewers should also note the category convention: the clinician remains responsible for the full study, and the software's output is an attention aid, not a screen-out.
Intended Use & Indications
Deep Capsule is a software-only reading aid for small-bowel capsule endoscopy. Its FDA-cleared indication (K250655, decided 12 March 2026) reads, in the 510(k) summary's wording, as an artificial-intelligence-assisted reading tool designed to aid small-bowel capsule-endoscopy reviewers in decreasing the time to review capsule-endoscopy images for adult patients in whom the images were obtained for suspected small-bowel bleeding, with the clinician responsible for conducting their own assessment through review of the entire video. The scope is deliberately limited. The same summary states that the device is capable of detecting small-bowel lesions without differentiating them, that it is not intended to characterise lesions in a manner that would replace biopsy or other characterisation tools, and that it is not a stand-alone diagnostic. Its AI output is a gallery of suspected-lesion frames selected from the video; a findings count and exam metadata are non-AI outputs. The device belongs to the FDA's young capsule-endoscopy analysis software category (product code QZF, 21 CFR 876.1540, established by the NaviCam ProScan De Novo), in which the clinician remains the reader of record. Compatibility is limited to the small-bowel capsule systems named in the labelling.